Repository logo
 
Loading...
Thumbnail Image
Publication

Chlormequat selective electrodes: construction, evaluation and application at FIA systems

Use this identifier to reference this record.
Name:Description:Size:Format: 
ART_GoretiSales_2003_GRAQ.pdf347.52 KBAdobe PDF Download

Advisor(s)

Abstract(s)

A flow injection analysis (FIA) system having a chlormequat selective electrode is proposed. Several electrodes with poly(vinyl chloride) based membranes were constructed for this purpose. Comparative characterization suggestedthe use of membrane with chlormequat tetraphenylborate and dibutylphthalate. On a single-line FIA set-up, operating with 1x10-2 mol L-1 ionic strength and 6.3 pH, calibration curves presented slopes of 53.6±0.4mV decade-1 within 5.0x10-6 and1.0x10-3 mol L-1, andsquaredcorrelation coefficients >0.9953. The detection limit was 2.2x10-6 mol L-1 and the repeatability equal to ±0.68mV (0.7%). A dual-channel FIA manifold was therefore constructed, enabling automatic attainment of previous ionic strength andpH conditions and thus eliminating sample preparation steps. Slopes of 45.5±0.2mV decade -1 along a concentration range of 8.0x10-6 to 1.0x10-3 mol L-1 with a repeatability ±0.4mV (0.69%) were obtained. Analyses of real samples were performed, and recovery gave results ranging from 96.6 to 101.1%.

Description

Keywords

Chlormequat Growth regulator Pesticides Potentiometry Ion-selective electrode FIA

Citation

Research Projects

Organizational Units

Journal Issue

Publisher

Taylor & Francis

CC License

Altmetrics