Long-term electrode–skin impedance variation for electromyographic measurements

dc.contributor.authorSousa, Andreia S. P.
dc.contributor.authorNoites, Andreia
dc.contributor.authorVilarinho, Rui
dc.contributor.authorSantos, Rubim
dc.date.accessioned2023-10-27T15:14:22Z
dc.date.available2023-10-27T15:14:22Z
dc.date.issued2023-10-19
dc.description.abstractThis study aims to observe the evolution of the electrode–skin interface impedance of surface EMG electrodes over the time taken to determine the time of stabilization. Eight healthy subjects participated in the study. Electrode–skin impedance was evaluated in the rectus abdominal muscle every five minutes, over a total period of 50 min. A reduction of 13.23% in the impedance values was observed in minute 10 (p = 0.007), and a reduction of 9.02% was observed in minute 15 (p = 0.029). No statistically significant differences were observed in the other instants evaluated. The findings obtained in the present study demonstrate a decrease in electrode–skin impedance from minute 5 to minute 15, followed by a stabilization period with a low percentage of variation till minute 50.pt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.citationSousa, A. S. P., Noites, A., Vilarinho, R., & Santos, R. (2023). Long-Term Electrode–Skin Impedance Variation for Electromyographic Measurements. Sensors, 23(20), Artigo 20. https://doi.org/10.3390/s23208582pt_PT
dc.identifier.doi10.3390/s23208582pt_PT
dc.identifier.eissn1424-8220
dc.identifier.urihttp://hdl.handle.net/10400.22/23803
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.publisherMDPIpt_PT
dc.relationThis research was supported by the Fundação para a Ciência e Tecnologia (FCT) through R&D Units funding [UIDB/05210/2020].pt_PT
dc.relation.publisherversionhttps://www.mdpi.com/1424-8220/23/20/8582pt_PT
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/pt_PT
dc.subjectElectrode–skin impedancept_PT
dc.subjectElectromyographypt_PT
dc.subjectStabilitypt_PT
dc.titleLong-term electrode–skin impedance variation for electromyographic measurementspt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.endPage6pt_PT
oaire.citation.startPage1pt_PT
oaire.citation.titleSensorspt_PT
oaire.citation.volume23 (20)pt_PT
person.familyNamePinheiro de Sousa
person.familyNameNoites
person.familyNameVilarinho
person.familyNameRubim Silva Santos
person.givenNameAndreia Sofia
person.givenNameAndreia
person.givenNameRui
person.givenNameManuel
person.identifier1070119
person.identifier1429505
person.identifier.ciencia-id2216-9200-7EF6
person.identifier.ciencia-id4918-EB69-ED9C
person.identifier.ciencia-idD711-9AB4-70AE
person.identifier.ciencia-idA018-2B24-C597
person.identifier.orcid0000-0001-9528-1463
person.identifier.orcid0000-0002-2590-962X
person.identifier.orcid0000-0002-4422-8440
person.identifier.orcid0000-0002-7394-7604
person.identifier.ridC-7138-2019
person.identifier.scopus-author-id55950021600
person.identifier.scopus-author-id55749467200
rcaap.rightsopenAccesspt_PT
rcaap.typearticlept_PT
relation.isAuthorOfPublicationaeecfe02-e80d-49ab-9033-a1ade15658f2
relation.isAuthorOfPublication1cd76bec-4c53-44eb-b484-5f2d774b7fbb
relation.isAuthorOfPublication04db8c5d-9644-460d-b318-dbad791becbd
relation.isAuthorOfPublication0a264044-a870-4fca-b976-f7ce003d23a8
relation.isAuthorOfPublication.latestForDiscovery1cd76bec-4c53-44eb-b484-5f2d774b7fbb

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
ART_Andreia Sousa.pdf
Size:
648.96 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: