Optimization and validation of [13N]-NH3 production for clinical studies of positron emission tomography in the evaluation of myocardial perfusion

Use this identifier to reference this record.
Name:Description:Size:Format: 
COM_LuisMetello_2015.pdf364.84 KBAdobe PDF Download

Advisor(s)

Abstract(s)

The present work aims to validate the [ 13 N]-NH 3 production for human studies of myocardial perfusion using a cyclotron from a Portuguese public institution and an automatic synthesis module. According to specifications of European Pharmacopeia (Ph. Eur.) 8.2., 3 consecutives batches were evaluated for this purpose. Results showed a clear and colorless product and pH values between 4,5 e 8,5. The chemical purity was confirmed by the colorimetric test and the samples were identified as [ 13 N]-NH 3 , by ionic chromatography, gamma ray spectroscopy and half-life determination. Radiochemical and radionuclidic purity were confirmed, corresponding a minimum of 99% of total radioactivity to [ 13 N] and a percentage of impurities less than 1%, 2 hours after the end of synthesis. Tests for sterility and for presence of bacterial endotoxins were also performed.