Repository logo
 
Publication

Flow injection potentiometric determination of chlorpromazine

dc.contributor.authorSales, M. Goreti F.
dc.contributor.authorTomás, José F. C.
dc.contributor.authorLavandeira, Sandra R.
dc.date.accessioned2013-11-29T11:13:22Z
dc.date.available2013-11-29T11:13:22Z
dc.date.issued2006
dc.description.abstractNew chlorpromazine selective electrodes with a tubular arrangement and no internal reference solution are proposed. Selective membranes are of poly(vinyl chloride) (PVC) with the tetraphenylborate•chlorpromazine (TPB•CPZ) ion-exchanger dissolved in o-nitrophenyl octyl ether (oNPOE). Analytical features of the electrodes were evaluated on a single-channel flow assembly having 500 µl injection volumes and flow-rates of 4.5 ml min−1. For a carrier solution of 3.3×10−3Min sodium sulphate, Nernstian responsewas observed over the concentration range 1.0×10−5 to 1.0×10−2 M. Average slopes were about 59mVdecade−1 and squared correlation coefficients were >0.9984. Slight hiper-Nernstian behaviour was observed in buffer solutions of 4.4 pH; average slopes were of 62.06mVdecade−1. The electrode displayed a good selectivity for CPZ, with respect to, several foreign inorganic and organic species. The selective electrodes were successfully applied to the analysis of pure solutions and pharmaceutical preparations. Proposed method allows the analysis of 84 samples h−1, producing wastewaters of low toxicity. The proposed method offers the advantage of simplicity, accuracy, applicability to coloured and turbid samples, and automation feasibility.por
dc.identifier10.1016/j.jpba.2006.03.027
dc.identifier.doi10.1016/j.jpba.2006.03.027pt_PT
dc.identifier.urihttp://hdl.handle.net/10400.22/3018
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherElsevierpor
dc.relation.ispartofseriesJournal of Pharmaceutical and Biomedical Analysis; Vol. 41, Issue 4
dc.relation.publisherversionhttp://www.sciencedirect.com/science/article/pii/S0731708506002767#por
dc.subjectChlorpromazinepor
dc.subjectIon-selective electrodespor
dc.subjectFlow injection analysispor
dc.subjectPotentiometrypor
dc.subjectPharmaceuticalspor
dc.titleFlow injection potentiometric determination of chlorpromazinepor
dc.typejournal article
dspace.entity.typePublication
oaire.citation.endPage1286por
oaire.citation.startPage1280por
oaire.citation.titleJournal of Pharmaceutical and Biomedical Analysispor
oaire.citation.volume41por
rcaap.rightsclosedAccesspor
rcaap.typearticlepor

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
ART_GoretiSales_2006_GRAQ.pdf
Size:
96.21 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: